7. what happens to the total number of x-ray interactions that occur as kvp is increased? why?

7. what happens to the total number of x-ray interactions that occur as kvp is increased? why?
7. what happens to the total number of x-ray interactions that occur as kvp is increased? why?

The images shown above were obtained at 50 kV (left) and 125 kV (right) at table top (i.e., with no scatter removal Bucky). The low kV image has superior contrast because of reduced levels of scatter. As the kV is reduced, the average photon energy is reduced; accordingly, the proportion of photoelectric interactions increases and the proportion of scatter events is reduced. Scatter is generally lower at lower kV values, especially when imaging bone which contains a high atomic number material (Calcium, Z = 20). In general, extremity imaging may be performed at table top because two factors help to reduce the amount of scatter: (a) thin body parts are easier to penetrate and therefore permit the use of lower x-ray tube voltages (kV); (b) the presence of bone (high Z) means that most interactions are photoelectric (not Compton) and thus scatter is reduced.

The image on the left (i.e., 50 kV), for a constant receptor radiation dose of 5 mGy (S ~ 200), required 15 mAs whereas the image on the right (125 kV) required only 0.5 mAs to achieve the same receptor dose. Increasing the x-ray tube voltage from 50 to 125 kV increases both the total x-ray tube output (air kerma) as well as the average photon energy (i.e., penetrating power). For both of these reasons, increasing the x-ray tube voltage by this amount required a 30 fold reduction in the mAs to maintain the same amount of radiation incident on the image receptor. The dynamic range of the image obtained at 50 kV was 160 (i.e., L = 2.2), whereas the dynamic range of the image obtained at 125 kV was markedly reduced to 50 (i.e., L = 1.7). In radiology, it is generally true that high x-ray tube voltages (i.e., increased photon energies) reduce the dynamic range in radiographic images, and vice versa.

7. what happens to the total number of x-ray interactions that occur as kvp is increased? why?
7. what happens to the total number of x-ray interactions that occur as kvp is increased? why?

The AP skull radiographs shown above were also obtained at the table top using 65 kV (left) and 125 kV (right). As expected, the low kV image is markedly superior to the high kV image because of the substantially lower levels of scatter. To achieve the same receptor dose, the mAs for the 125 kV image on the right was reduced by a factor of about 15 - the 125 kV image was obtained using 0.66 mAs, whereas the 65 kV image was obtained using 10 mAs, and where both resulted in an image receptor radiation intensity of 5 uR (i.e., S ~ 200).

7. what happens to the total number of x-ray interactions that occur as kvp is increased? why?
7. what happens to the total number of x-ray interactions that occur as kvp is increased? why?

The lateral skull radiographs shown above were also obtained at the table top using 60 kV (left) and 125 kV (right). Note that one can use a lower kV (60 kV) on the lateral skull than the AP (65 kV) because the lateral is generally thinner and therefore easier to penetrate. As expected, the low kV image is markedly superior to the high kV image because of the substantially lower levels of scatter.To achieve the same receptor dose, the mAs for the 125 kV image on the right was reduced by a factor of about ~25 - the 125 kV image was obtained using 0.5 mAs, whereas the 60 kV image was obtained using 12 mAs, with both resulting in an image receptor radiation intensity of 10 mR (i.e., S ~ 100).






Scatter radiation, introduced in Chapter 2, is produced as a result of the attenuation of the x-ray beam by matter. This chapter explores the production of scatter radiation and the factors that influence its formation. In addition, this chapter covers methods used to minimize the fog that this radiation causes on radiographs.




When x-rays completely penetrate the body, there is no interaction with matter, and no scatter or scattered radiation is formed as a result. When x-rays are absorbed in the body, however, their energy is “scattered,” or converted into new scatter x-rays. Three types of interactions occur when radiation is absorbed by matter: coherent scattering, Compton effect, and photoelectric effect.


The result of either coherent scattering or the Compton effect is termed scatter radiation or simply scatter. Radiation produced by the photoelectric effect is correctly referred to as secondary radiation. Since more than one type of interaction takes place during radiography and the resulting radiation is so similar, the terms are often used interchangeably. When referring to both scatter and secondary radiation, this text uses the term scatter radiation.


The interactions that produce scatter radiation in radiography occur primarily within the patient. Some scattering also occurs as a result of interactions between the x-ray beam and the tabletop and image receptor (IR), and any other matter that happens to be within the radiation field.






The photoelectric effect is similar to that which forms characteristic radiation in the x-ray tube (see Chapter 5). In this case, however, the incoming energy is an x-ray photon interacting with an atom in the body rather than an electron interacting with the tungsten anode.


In a photoelectric interaction, the incoming photon from the primary beam collides with an inner orbital electron of an atom. The photon is totally absorbed in the process and creates an absorbed dose in the patient. The electron’s departure leaves a “hole” in the orbit, which is filled by an electron from an outer shell. The difference in binding energy between the two shells is emitted as a new x-ray photon (Fig. 9-3). This photon is referred to as a characteristic photon and is considered secondary radiation because it is radiation actually produced in the body. The photon will have a new direction. Its energy will be less than that of the primary photon. Photoelectric interactions are less prevalent in the diagnostic energy range than Compton interactions. The likelihood of a photoelectric interaction is determined by both the kVp level and the electron-binding energy of the atom in which the interaction occurs.



Because no part of the energy of the incoming photon exits the atom, photoelectric interactions are sometimes referred to as true absorption. In this text, references to scatter also apply to secondary radiation formed by the photoelectric effect. As kVp is increased, photoelectric effect is decreased. Note, this is the opposite of the Compton effect. In the diagnostic range of kVp used (50 to 100 kVp) the majority of radiation interactions with the body are Compton interactions.



Conventional X-ray production involves the excitation of tungsten metal to release photons.[1][2][3][4] A cathode ray is used to direct energy into the rotating tungsten filament anode. The resultant photons released can be absorbed or transmitted through the body to provide information on the amount of attenuation. These attenuation gradients are used to reconstruct an image by mapping the amount of primary or returning photons that hit a photosensitive detector plate and produce a planar image.[5][6] Other fates of photons may include scattering, in which photons are deflected away from the detector.[7] 

Alternatively, photons can be completely absorbed into the tissue.[8] Physical tissue properties may determine whether energy is more easily absorbed, attenuated, or scattered. In particular, tissue density, thickness, and atomic number alter the trajectory and absorption of X-rays.[9][10] Increased atomic number, thickness, and density can cause photons to be attenuated, absorbed, and scattered to higher degrees. These properties create contrast among different tissues within the body, allowing for a separation of intensity values and evaluating potential pathology.

X-ray image production procedures focus on optimizing settings to produce the appropriate contrast among the anatomy of interest while limiting noise and artifacts that may detract from the evaluation of the image.[3] It is often a trade-off to optimize imaging parameters while keeping ionizing radiation exposure as low as reasonably achievable (ALARA). Important aspects of determining appropriate protocol in clinical settings include X-ray tube voltage (kVp), current (mA), and exposure time (seconds).[11][12][13][14]

X-ray Production

kVp

The kilovoltage peak (kVp) is the difference in potential applied to the X-ray tube.[11][14] kVp is directly proportional to the average energy of the X-ray spectrum produced, referred to as X-ray quality.[14] kVp plays a role in adjusting the amount of penetration and exposure in an acquisition. Penetrance is characterized by the number of photons reaching the image receptor to discern differences between structures. For example, in underexposed chest X-ray acquisitions in which the diaphragm cannot be visualized to the intersection of the spine, kVp can be increased to mitigate this issue.  An adequate penetrance ensures the ability to separate definable structures of interest; recent advances have allowed altering digital windowing levels to achieve the same effect. Changes in kVp affect radiation dose, exposure, and contrast. Also, dose increases proportionally with higher kVp. Exposure doubles in intensity for every 15% increase in kVp, whereas contrast decreases with increases in kVp.[11] 

In contrast, this decrease is primarily due to an increased proportion of Compton scatter at higher kVp.[7] Compton scatter is one of two primary methods in which X-rays interact with matter, the other being the photoelectric effect. An increased ratio of Compton scatter introduces excess photons that reach the image receptor. As a result, the image becomes overexposed. To achieve the best possible results, kVp is increased for sufficient exposure but kept low to minimize overexposure and radiation dose.

mAs

Milliamperes (mA) is a unit representing the amount of current passed through the X-ray tube. Current determines the number of photons produced by the X-ray tube, also known as X-ray quantity.[12] Another contributing factor toward X-ray quantity is the total exposure time, measured in seconds. Current and exposure time are often reported together such that: current(mA) x time(s) = milliampere-seconds (mAs). Changes in mAs affect radiation dose, signal-to-noise ratio (SNR), and contrast.[14] Increasing mAs produces more electrons in an X-ray tube and subsequently increases the amount of radiation exposure.[11] High mAs will increase SNR but will decrease image contrast. X-ray imaging protocols are designed to optimize SNR while maintaining adequate contrast and limiting radiation dose.

Factors that Influence Image Quality

Contrast

Effectively determining anatomy and suspected pathologies rely on identifying and separating different tissue types and boundaries. In X-ray imaging, contrast describes the number of relative photons that can pass through a tissue comparative to another. This is determined by the amount of tube voltage (kVp) and filtration used. Conversely, increasing the mA does not improve or worsen contrast and contributes to the amount of noise in the image.[14] Selecting parameters with lower kVp will allow for the best separation in a given spectrum of intensities and consequently improve contrast. However, this is always balanced with achieving enough exposure and penetration. Another technique utilized to improve contrast to noise is using a grid to reduce scatter.[9] The choice of the grid is based on imaging modality (breast, abdomen, skull) and grid spacing ratios contribute to the amount of noise reduction. 

Distortion

Beam profiles and paths of the photons also influence the quality and characteristics of an image. X-ray divergence patterns can be described by photons directed linearly towards the center, whereas those on the periphery tend to splay out more radially.[8] As a result, the anatomy located on the periphery of the beam profile and lateral to the center will suffer some degree of distortion. However, some commonly manipulatable factors can limit the amount of distortion in an image: centering, source image receptor distance (SID), and object image receptor distance (OID). Centering refers to positioning the anatomical portion of interest in line with the central point of the X-ray. SID is the distance between the X-ray tube and the image receptor and is inversely proportional to magnification/distortion. In other words, the greater the SID, the less magnification/distortion will be apparent in the image. The standard SID (ref) utilized is set to be 100 cm. OID is the distance between the object (e.g., femur, abdomen) and the image receptor, which is directly proportional to magnification. The greater the OID, the greater the magnification. Taking all three factors into account, the most optimal positioning for X-ray imaging would be to have the anatomy of interest in the center of the X-ray beam, the beam sufficiently distanced from the image receptor, and the image receptor as close as possible to the anatomy being imaged.[15]

Mottle

Also known as quantum noise, mottle is noise due to random distribution and an uneven number of photons reaching the image detector.[16] Mottle is the largest contributor to noise in plain X-ray; it is mostly a consequence of images acquired with low radiation doses.[14] The noise generates graininess in an image, thus disrupting the uniformity of the image. Mottle can be reduced by using a higher mA, which will increase the average number of photons and the SNR.[12]

Spatial Resolution

Another factor considered in image quality is the spatial resolution, which is determined by measuring the smallest distinguishable space between two distinct lines or landmarks. The smaller the distance between line pairs relates to discerning boundaries and colloquially defined sharper or better resolution images. One of the changeable features that may influence spatial resolution is anode angle. Anode angle is the relationship between the slant of the tungsten anode and the incident cathode ray. The degree of the anode angle significantly contributes to the size of the focal spot generated. Lower amounts of anode angle relate to a smaller focus and image with better spatial resolution.[11]

Beam Filtration

Beam filtration refers to the use of X-ray absorbing material (e.g., copper, aluminum, titanium) placed between the X-ray beam and patient to increase the average photon energy by absorbing lower energy photons.[17] These low energy photons detract from image quality by increasing the amount of scatter and unnecessarily contributing to increased patient dose.[11] Filtration reduces  Compton scatter and has the effect of decreasing X-ray quantity and increasing X-ray quality. The clinical effects of beam filtration include increased image contrast at the cost of increased patient exposure.[17]

Grid

Scatter reduction is primarily addressed with the use of grids. A grid is placed between the patient and the receptor and is composed of X-ray absorbing material (e.g., lead) interspaced with low attenuating material (e.g., carbon fiber).[11] The amount of scatter reduction a grid provides is directly proportional to the ratio between the height of the grid and the interspacing, also known as the grid ratio. The greater the grid ratio (e.g., 10:1, 12:1), the greater the amount of scatter reduction, which also increases image contrast and patient exposure dose. The ratio of increase in image contrast and the patient dose is referred to as the contrast improvement factor and the Bucky factor, respectively.[14]

Anode Heel Effect

The anode heel effect describes the phenomenon of the gradient of X-ray emission relative to the angle of X-ray toward the cathode. The number of X-rays emitted is inversely proportional to the angle of emission relative to the cathode.[11] This difference in photon production is a consequence of tungsten excitation beneath the surface of the anode. The X-rays produced within the anode must travel through the material before being emitted. As a result, fewer X-rays are produced in areas where more material must be traversed. X-rays are produced in a gradient, with the highest beam strengths found closest to the cathode.[12] This effect is more pronounced at lower anode angles; angles < 6o are not recommended in clinical practice due to this phenomenon.[14]

X-ray image production procedures utilize a balance of image optimization, contrast, distortion, noise, and patient dose. However, these are only a few of many factors technicians and radiologists must consider optimizing when scanning patients and selecting appropriate protocols. Often these settings offer improvement of one imaging parameter at the cost of another. For example, devices (e.g., filters, grids) enhance image contrast at the cost of increased patient dose. Understanding these parameters is central to the understanding and use of X-ray imaging in diagnostic medicine. These understandings demonstrate further utility with interventionalists in radiology, surgery, and pain medicine who use X-ray imaging in real-time to guide therapeutic interventions.